PERFUSION SYSTEMS BIO CONSOLE 560; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 560BCS1
Device Problems Electrical Power Problem (2925); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/28/2019
Event Type  Injury  
Manufacturer Narrative
Product has not been returned for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during use, this bio-console instrument displayed an "ac power failure" error when connected to ac power.There was no reported malfunction after the error was displayed and the instrument was functioning.The hand-crank was used temporarily while replacing the instrument with a backup.It is asked but unknown if there was any effect on the patient.
 
Manufacturer Narrative
The bio-console was returned and the reported "ac power failure" error message was verified during service.The issue was resolved by replacing the power supply circuit board.Preventive maintenance was completed per specifications.The batteries and fan filters were replaced as part of the preventive maintenance.Trends for this with this product are monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO CONSOLE 560
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
MDR Report Key8788210
MDR Text Key150956313
Report Number2184009-2019-00046
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K080824
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number560BCS1
Device Catalogue Number560BCS1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2019
Date Manufacturer Received08/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-