| Catalog Number |
6020-0637 |
| Medical Device Problem Codes |
Degraded (1153); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Pain (1994); Injury (2348); Reaction (2414); Metal Related Pathology (4530)
|
| Date of Event |
05/14/2010
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.Not returned.
|
| |
|
Event or Problem Description
|
|
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2008 and was revised on (b)(6) 2010.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue and excessive levels of chromium and cobalt in his blood.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding abnormal ion level involving an accolade stem was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical records by a clinical consultant indicated: "the implant record and operative report confirm a revision tha was performed for loss of acetabular fixation with change in component position in a cemented tha.No documentation was provided regarding increased metal ion levels." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to elevated metal ion levels.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2008 and was revised on (b)(6) 2010.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue and excessive levels of chromium and cobalt in his blood.As per medical review: "the implant record and operative report confirm a revision tha was performed for loss of acetabular fixation with change in component position in a cemented tha.No documentation was provided regarding increased metal ion levels." medical record: "[.] followup radiograph of the hip was obtained and there was obvious change in position of the acetabular component with loss of fixation of the cement.
|
| |
|