| Model Number |
7010 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast in the inflation lumen.The balloon was tightly folded.There were numerous hypotube kinks throughout the device.There was a complete hypotube separation 50cm distal of the strain relief.The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation.The tip was damaged.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 24jun2019.It was reported that shaft kink occurred.The target lesion was located in the left coronary artery.A 2.5mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, it was noted that the delivery shaft was kinked.The procedure was completed with another of the same device.No patient complications were reported that the patient's status was stable.However, returned device analysis revealed separated patient's.
|
| |
|