ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and evaluated by product analysis on 12-jul-2019 the following findings: during investigation, the battery compartment was found to be cracked.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(6).
 
Event or Problem Description
The pump was returned for investigation.Investigation revealed a battery compartment crack.This report is made based on results of investigation completed on 12-jul-2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key8790054
Report Number2531779-2019-04071
Device Sequence Number1904508
Product Code OYC
UDI-Device Identifier10840406100051
UDI-Public10840406100051
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2019
Was the Report Sent to FDA? Yes
Device Age49 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2019
Initial Report FDA Received Date07/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight200
-
-