Device Problems Break (1069); Entrapment of Device (1212)
Patient Problem Foreign Body In Patient (2687)
Event Date 05/29/2019
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation and no images or videos have been provided of the filter in-vivo, so the complaint cannot be confirmed.If additional information is provided in the future, this issue will be reevaluated as needed.
 
Event Description
Subject is a (b)(6) year old female enrolled in preserve on (b)(6) 2017 with option¿ elite retrievable vena cava filter placed on same day due to history of right lower extremity dvt and contraindication to anticoagulation due to upcoming embolization of basilar apex aneurysm surgery.Subject was discharged from hospital on (b)(6) 2017.During filter retrieval on (b)(6) 2019 site reported that after snaring the hook of the filter, it was noted the single left sided filter leg remained stationary while the rest of the filter removed freely on the snare, making it clear that this single leg was fractured from the rest of the filter.After successful removal of the filter, the embedded leg was then successfully retrieved using the snare with no further fracturing.The event was considered definitely related to the ivc filter or procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTION IVC FILTER
Type of Device
RETRIEVABLE INFERIOR VENA CAVA FILTER
Manufacturer (Section D)
ARGON MEDICAL DEVICES INC.
1445 flat creek road
athens TX 75751
Manufacturer (Section G)
ARGON MEDICAL DEVICES INC.
Manufacturer Contact
gail smith
1445 flat creek road
athens, TX 75751
2144368995
MDR Report Key8790844
MDR Text Key151747736
Report Number1625425-2019-00193
Device Sequence Number1
Product Code DTK
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K133243
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type study,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/01/2005,06/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 06/26/2019
Initial Date FDA Received07/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-