MDT POWERED SURGICAL SOLUTIONS LEGEND DISSECTING TOOL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number 14CY50BDC
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Perforation (2001)
Date of Event 06/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the device was discarded by the customer.The device user manual warnings section includes instructions to check the device for damage before use.If damage is found, the device should not be used.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the dura was damaged because the tip was hard to see and depth was difficult to know.No further information can be provided on the current status of the patient.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received.Initially, cervical artificial disc replacement was planned to be performed, but it was changed to anterior cervical corpectomy and fusion.It was confirmed that there was a 45-minute delay on the procedure to stop the cerebrospinal fluid leak caused by dural injury.The procedure was completed with backup product.Further information about the delay is being followed up from the facility contact person.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
On follow up, it was confirmed that there was no unplanned procedure and the delay was due to preventing the cerebrospinal fluid to leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND DISSECTING TOOL
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key8790985
Report Number1625507-2019-00069
Device Sequence Number1977300
Product Code HBE
UDI-Device Identifier00643169760257
UDI-Public00643169760257
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2022
Device Model Number14CY50BDC
Device Catalogue Number14CY50BDC
Device Lot Number0214445718
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2019
Supplement Date Received by Manufacturer07/16/2019
11/11/2019
Initial Report FDA Received Date07/15/2019
Supplement Report FDA Received Date08/14/2019
12/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-