(b)(4).Device evaluation summary: one detached needle and one suture piece of product code vcp364, lot mjm737 were returned for analysis.During the visual inspection of the detached needle, the swage and attachment area were noted to be as expected, but marks appears to be by use of surgical instrument could be found on the body needle.The suture piece was examined.A correct insertion mark and body fluids were observed; also, the end was cut.In addition, the force used to detach needle from the suture could not be determined.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the sample condition, it could not be determined what may have caused the reported incident of ¿2 threads were found detached from the needles of vcp3364h during a caesarean section.¿ a manufacturing record evaluation was performed for the finished device mjm737 number, and no non-conformances were identified.Expiration date: 07/31/2023.Date of mfg.: 08/07/2018.A manufacturing record evaluation was performed for the finished device mjm737 number, and no non-conformances were identified.First device reported in mw 2210968-2019-84292.
|