| Model Number |
S5-019 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
05/17/2019
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Device location not presently known.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, a top hat size 19 was reportedly implanted/explanted on the same day.Based on the information reported, the patient passed away on the same day.No other information is available at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, a top hat size 19 was reportedly implanted/explanted on the same day.The device did not fit in the patient's annulus and a freestyle root was ultimately implanted in the patient.There was no perceived issue with the explanted top hat and the event was attributed to the patient's small annulus.The patient passed away on the same day, but per the surgeon's assessment, the device was not suspected to be a contributory cause.
|
| |
|
Additional Manufacturer Narrative
|
|
A complete manufacturing and material records review for the device has been performed.The results confirmed that the device satisfied all material, visual and performance standards required at the time of manufacture and release.Based on the site's assessment, the device explant occurred due to mis-sizing due to the patient's small annulus.As such, the root cause of the adverse event is deemed to be related to patient's factors and not to the device.Regarding the patient's death, it was confirmed that it occurred after the top hat explant and it was deemed not device-related.
|
| |
|