| Model Number |
401769 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
After direct stenting the orsiro could not pass the severely calcified lesion in the rca and was therefore removed from the patient's body.During removal the stent dislodged from the balloon.Eventually the dislodged stent was covered and tacked into place with another balloon.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor the angiographic material was provided for analysis.Therefore no technical investigation on the subject could be performed.The manufacturing history was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the manufacturing history of the product detailed above did not reveal any non conformity.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.The device in question was manufactured according to specifications and successfully passed all inprocess controls as well as the final inspection.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.It should be noted that according to the ifu predilatation of the lesion is mandatory.
|
| |
|