Model Number 401735
Medical Device Problem Codes Failure to Advance (2524); Device Dislodged or Dislocated (2923); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/06/2019
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting stent system was chosen to treat a lesion (100 percent stenosis degree) in a very tortuous rca.After pre-dialtation the device could not cross lesion.After withdrawal the stent dislodged outside patient's body and appeared to be deformed.
 
Additional Manufacturer Narrative
The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.Both the delivery system and the dislodged stent were returned for analysis.The technical investigation of the returned device revealed that the balloon is still well folded and does not show any signs of inflation.Microscopic analysis of the balloon surface revealed clear visible stent imprints, indicating that the stent was properly crimped in between the two radiopaque markers at the time of delivery.The dislodged stent was returned deformed at one end.Review of the manufacturing history of the product detailed above did not reveal any non conformity.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.The device in question was manufactured according to specifications and successfully passed all inprocess controls as well as the final inspection.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.The final root cause is most likely related to the handling during the procedure.It should be noted that the ifu warns against reinsertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.25/13
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key8791644
Report Number1028232-2019-02904
Device Sequence Number1822583
Product Code NIQ
UDI-Device Identifier07640130438903
UDI-Public07640130438903
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number401735
Device Catalogue NumberSEE MODEL NO.
Device Lot Number09184428
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2019
Initial Date Received by Manufacturer 06/27/2019
Supplement Date Received by Manufacturer09/04/2019
Initial Report FDA Received Date07/15/2019
Supplement Report FDA Received Date09/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-