• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PROLENE SUTURE 30IN(75CM) 2-0 BLU; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PROLENE SUTURE 30IN(75CM) 2-0 BLU; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number W295H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint (b)(4).To date the device has not been returned.If the device or further details are received at the later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: what procedures were being completed? hernia repair under local anesthetic.What part of the procedure was impacted? none all procedures completed with another batch of sutures.Are all 3 devices had issues in these single procedures? not all separate procedures.How was case completed? with a new box of sutures with different batch number.Any patient consequences? patient had to be sent for x-ray to ensure tip was not present in body.Received conformation (b)(6) 2019 from x-ray, no tip visible.What medical/surgical intervention was provided to manage them? no medical or surgical intervention needed.Did 2 devices of w295h had needle breakage in this procedure? if no, please explain.Needle tips both broke off while suturing in mesh during hernia repair.Was there an impact on the patient? if so, was there an increase in procedure time form the suture issues? yes, as tip of prolene on the first patient could not be located so patient sent for x-ray to confirm not in patient waiting on results second patient tip was located and retrieved.
 
Event or Problem Description
It was reported that the patient underwent a hernia repair procedure in 2019 under local anesthetic and the suture was used.During the procedure, a needle tip broke off while suturing in mesh during hernia repair and could not be located.Another like suture was used to complete the procedure.The tip was located and retrieved there were no patient consequences reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE SUTURE 30IN(75CM) 2-0 BLU
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key8792110
Report Number2210968-2019-84310
Device Sequence Number16229160
Product Code GAW
UDI-Device Identifier10705031238244
UDI-Public10705031238244
Combination Product (Y/N)N
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Catalogue NumberW295H
Device Lot NumberMPH640
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/19/2019
Supplement Date Received by Manufacturer08/26/2019
Initial Report FDA Received Date07/16/2019
Supplement Report FDA Received Date09/23/2019
Patient Sequence Number1
-
-