| Catalog Number |
K13A |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent balloon kyphoplasty surgery due to primary osteoporosis.Intra-op, the balloon did not deflate even though the handle was pulled while the trigger of the inflatable bone tamp (ibt ) was being gripped.Since the balloon would deflate when having the handle rotated, the operation was completed without problems as the balloon deflated when the handle was rotated.There was a delay of less than 60 min in overall procedure time as a result of this event.There were no patient symptoms or complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis results: visual and functional inspection confirmed the syringe handle is not working properly.The spring has become dislodged which is keeping the handle jammed in place.The handle may have been pulled back with excessive force causing the spring to pop out.Root cause of non-conformance is undetermined.If information is provided in the future, a supplemental report will be issued.
|
| |
|