| Model Number |
MMT-551NAL |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/06/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had blank display.The customer¿s blood glucose level was 140 mg/dl at the time of the incident.Customer stated the contacts on the battery cap were not missing or damaged.Customer stated the battery compartment was neither damaged nor corroded.Customer stated the spring was neither damaged nor corroded.Customer stated that the display did not returned.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device monitored for several days and no blank display noted.Device received with normal operating current.Device passed self test.Device passed off no power test.(b)(4).
|
| |
|