|
This report has been identified as b.Braun medical internal report number (b)(4).The actual defective device was not received for evaluation.Retain samples for catalog 490438, lot 0061652783 were visual and functionally tested.Based on the evaluation results,1 out of 6 retain samples were found with leakage.The reported defect of leakage was confirmed.Per the manufacturer's statement, the defect is due to insufficent glue during the assembly process.In response to this incident and other incidents of this nature, the manufacturer has added new visual and written instructions to their procedure to ensure proper assembly.As a result of the retain failure for leakage, the house retain samples were also visually and fucntinoally tested for the before lot 0061649378 and after lot 0061658664.Lot 0061649378 was found acceptable; however, for lot 0061658664, 1 out of 5 units was found to have foreign matter/particulate in the fluid path.Please note, that mfr report number 9614279-2019-00191 will be filed for the confirmed defect of "foreign matter/particulate." furthermore, a review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|