• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US GII DRBIT 2.4MM *EA; SURGICAL HAND DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US GII DRBIT 2.4MM *EA; SURGICAL HAND DRILL Back to Search Results
Catalog Number 211093
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.Further, no lot numbers were supplied which precludes conducting a dhr review or a lot specific search in the complaints handling system.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).The lot number is unknown.
 
Event or Problem Description
It was reported by the sales rep via phone that during an elbow procedure it was found that the tip of their drill bit gii 2.4mm is dull.The procedure was completed with another like device with no patient harm or surgical delay to the case.The sales rep was not present for the case therefore could not provide any further information or a lot number for the device as it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SURGICAL HAND DRILL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8793614
Report Number1221934-2019-57672
Device Sequence Number2043767
Product Code HTW
UDI-Device Identifier10886705001576
UDI-Public10886705001576
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number211093
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/17/2019
Initial Report FDA Received Date07/16/2019
Patient Sequence Number1
-
-