Model Number G179
Device Problems High impedance (1291); Over-Sensing (1438); Failure to Select Signal (1582)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/28/2019
Event Type  Injury  
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined this device exhibited noise generated by the minute ventilation (mv)/respiratory sensor that is related to a high impedance condition.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) oversensed the minute ventilation (mv) sensor which was programmed off and successfully resolved the issue.Since then, this crt-d detected intermittent high out-of-range right atrial (ra) lead impedance measurements and decreased p-wave amplitudes.Boston scientific technical services (ts) recommended additional troubleshooting to rule out lead issues and/or lead connection issues.No adverse patient effects were reported.The device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOGEN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CLONMEL LIMITED
cashel road
,
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
,
saint paul, MN 55112
6515826168
MDR Report Key8794047
MDR Text Key151153429
Report Number2124215-2019-13603
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/12/2016
Device Model NumberG179
Device Catalogue NumberG179
Device Lot Number128248
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/23/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2014
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age74 YR
-
-