Catalog Number 284002
Device Problem Suction Failure (4039)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/17/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint : (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event Description
It was reported by the sales rep via phone that the outflow function of the fms vue pump was not working properly which is causing the loss of more saline than normal.They were able to complete the case with another like device with no surgical delay or no patient harm.This is all the information the sales rep has at this time.
 
Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received at the service center.The customer reported an issue of outflow was not working properly, per service manual operational and diagnostic analysis did not confirm the reported issue, therefore this complaint cannot be confirmed.The repair and testing of the unit will not be completed at this time because there is no need for it in the loaner equipment pool due to excess units in the pool at this time.Unit placed into long term hold.As no defect was identified, a root cause for the issue that was experienced by the user cannot be determined.A manufacturing record evaluation was performed for the finished device serial number, and no non-conformances related to the reported complaint condition were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key8794414
MDR Text Key151174013
Report Number1221934-2019-57677
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup,Followup
Report Date 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number284002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2019
Initial Date Manufacturer Received 06/18/2019
Initial Date FDA Received07/16/2019
Supplement Dates Manufacturer Received07/25/2019
08/02/2019
Supplement Dates FDA Received07/30/2019
08/07/2019
Patient Sequence Number1
-
-