| Model Number |
GM-322 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during an open hydrocelectomy, the device thread broke.The surgeon then used another suture to resolve the issue in order to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample noted one needle detached.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.Further microscopic inspection of the suture transfer material was noted inside the swage of the needle.As no sealed samples were returned, a needle attachment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.A definitive root cause could not be determined with regard to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|