MAUDE Adverse Event Report: NEODENT - JJGC S.A. CM TITAMAX EX ACQUA IMPLANT 3.5X15; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. CM TITAMAX EX ACQUA IMPLANT 3.5X15; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 140.664
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 02/18/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.The information provided in this report was based on information given by the dentist in neodent¿s products warranty form that was recorded in the system that is used by both the manufacturer and the subsidiary.The original complaint and the product were received by the subsidiary and did not arrive in the manufacturer yet.When this happens, a new evaluation of the complaint will be carried out and, if necessary, a follow-up report will be send.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.
 
Event or Problem Description
(b)(4) - the dentist reported that, after the dental implant was installed in intraoral region #6, its non-osseointegration was observed.The patient presented bone type ii.
 
Additional Manufacturer Narrative
The complaint was received by manufacturer and, after analysis, new informations were obtained.A follow-up is being sent to correct these informations according to the evaluation made.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.Moreover, it was seen that the dentist has used sequence of drills different than the one recommended for the implant installation.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form, visual conference of the product returned by the dentist and the evaluation of relevant production records.
 
Event or Problem Description
(b)(4).The dentist reported that, after the dental implant was installed in intraoral region #6, its non-osseointegration was observed.The patient presented bone type ii.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM TITAMAX EX ACQUA IMPLANT 3.5X15
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key8794948
Report Number3008261720-2019-03652
Device Sequence Number2638974
Product Code DZE
UDI-Device Identifier07899878016762
UDI-Public07899878016762
Combination Product (Y/N)N
PMA/510(K) Number
K133592
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 07/16/2019
Report Date (Section F) 06/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/21/2019
Device Catalogue Number140.664
Device Lot Number91100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/31/2019
Device Age25 MO
Event Location Other
Date Report to Manufacturer06/21/2019
Initial Date Received by Manufacturer 06/21/2019
Supplement Date Received by Manufacturer10/27/2019
Initial Report FDA Received Date07/16/2019
Supplement Report FDA Received Date10/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
-
-