ZIMMER BIOMET, INC. M2A-MAGNUM 42-50 TPR INSRT STD; PROSTHESIS HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Remove (1528)
Patient Problem No Information (3190)
Event Date 01/17/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical devices: m2a-magnum mod hd # item 157444 lot 012410.Ref 103204 lot 060690 taperloc femoral 10x140mm.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2019-03064.It is unknown if product will be returning to zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a revision surgery, the femoral head could not be released from the taper neck, leading to revision of the well-fixed femoral stem.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Medical records were received, however the event could not be confirmed.Medical records noted stem was well fixed.After multiple attempts, the stem was able to be removed.Dhr was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M2A-MAGNUM 42-50 TPR INSRT STD
Type of Device
PROSTHESIS HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8796021
MDR Text Key151221053
Report Number0001825034-2019-03065
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K042037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2019
Device Model NumberN/A
Device Catalogue Number139256
Device Lot Number916710
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-