| Model Number |
MMT-523LNAB |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Obstruction of Flow (2423); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that insulin pump buttons were difficult to press sometimes.Customer¿s blood glucose was unknown at the time of incident.The customer also stated that insulin pump had no delivery alarms and purple discoloration where cannula connects to the insulin pump.Troubleshooting was not performed.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
All buttons functioning properly.No button error alarm during testing.No moisture or corroded keypad found.No flatted keypad.Connector was inspected and locked on lcd board.Device passed displacement test, self test, off no power test and all operating currents tested within the specific range.No unexpected low battery alarm were noted.Device passed rewind test, basic occlusion test, prime/a33 test, excessive no delivery test and occlusion test, no unexpected no delivery alarm noted.(b)(4).
|
| |
|