| Catalog Number |
288314000 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination of the torque driver did not feature any immediately observable defects.The driver was mechanically tested and was found to be out of specification.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.A definitive root cause for the torque driver being out of calibration cannot be positively determined.Although not proven, the most probable root cause for the over torquing of the driver can be attributed to lack of maintenance during its time in use.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, during routine incoming receiving inspection, 6.0 nm torque limiting handle was found to be out of specification.There was no patient involvement.
|
| |
|