| Catalog Number |
0113320 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
06/26/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The subject product was discarded and not returned for evaluation.As reported, the physician's assistant placed their finger directly over the area where the fastener was deployed.The ifu for the optifix open fixation device precautions: counter pressure should be applied on the target area.Avoid placing hand/finger directly over the area where fastener is being deployed to prevent injury.Based on the information provided, the user did not follow the recommended instructions which caused the reported problem.A review of the manufacturing records was performed and found that the lot was manufactured to specification.Not returned.
|
| |
|
Event or Problem Description
|
|
It was reported by the bd sales rep.That during an open umbilical hernia repair, the surgeon was using the optifix open fixation device for the first time.The sales rep.Advised the surgeon to use adequate counter pressure.The surgeon's physician's assistant applied counter pressure on top on the fascia and under the skin and subq fat.At the first fire, the physician's assistant was hit in the finger and through her glove with the guide wire.The case was finished and the physician's assistant had her blood drawn.
|
| |
|