• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA¿ QUAD S CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA¿ QUAD S CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBB1QQ
Medical Device Problem Codes Failure to Capture (1081); Pacing Problem (1439)
Health Effect - Clinical Code Non specific EKG/ECG Changes (1817)
Date of Event 06/25/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: 6935m55 lead, implanted: (b)(6) 2019 459888 lead, implanted: (b)(6) 2019.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that there was loss of capture and pauses noted on a telemetry strip.Follow up concluded this was due to variable right ventricular (rv) lead thresholds and the device feature that monitors the pacing amplitude threshold and adjusts rv lead outputs.The device was reprogrammed and explanted and replaced.The rv lead was reprogrammed, revised, and remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was further reported that the patient was admitted for seizure like activity and long period of pauses was observed on the telemetry.The long period of pauses and loss of capture was due to the device feature that monitors the pacing amplitude threshold and adjusts rv lead outputs.It was noted that there was a possible micodislodgement due to the varying threshold.During the rv lead revision it was noted the rv lead had adhered to a previously implanted lead.There was also a left subclavian occlusion noted.The patient will be scheduled for a lead extraction.It was also noted that the physician elected to replace the cardiac resynchronization therapy defibrillator (crt-d) because they felt there should have been rv safety pacing when loss of capture occurs.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that there was loss of capture and pauses noted on a telemetry strip.Follow up concluded this was due to variable right ventricular (rv) lead thresholds, the rv lead output was possibly below the threshold, and the device feature that monitors the pacing amplitude threshold and adjusts rv lead outputs.The device was reprogrammed and explanted and replaced.The rv lead was reprogrammed, revised, and remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA¿ QUAD S CRT-D
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key8798401
Report Number3004209178-2019-13699
Device Sequence Number14074667
Product Code NIK
UDI-Device Identifier00643169720145
UDI-Public00643169720145
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2020
Device Model NumberDTBB1QQ
Device Catalogue NumberDTBB1QQ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2019
Initial Date Received by Manufacturer 07/17/2019
Supplement Date Received by Manufacturer07/17/2019
07/15/2019
08/01/2019
Initial Report FDA Received Date07/17/2019
Supplement Report FDA Received Date07/17/2019
07/24/2019
08/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1388TC46 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age88 YR
-
-