Model Number M0063301080
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a gemini stone retrieval basket was to be used for a transurethral lithotripsy procedure performed on (b)(6) 2019.According to the complainant, the device was unpacked, and a long hair was found inside the sterile packaging.The procedure was completed with another gemini basket.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a gemini stone retrieval basket was to be used for a transurethral lithotripsy procedure performed on (b)(6) 2019.According to the complainant, the device was unpacked and a long hair was found inside the sterile packaging.The procedure was completed with another gemini basket.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: device code 2944 captures the reportable event of foreign material present in device.Block h10: visual analysis found the pouch was completely opened and the device was out of its pouch when received.A hair was observed inside the plastic tray.The device was received out of its pouch which indicates handling and manipulation of the device.Due to this reason, it is unknown if the hair found in the plastic tray occurred during the manufacturing process or if it occurred as a result of manipulation of the device after it left the manufacturing site.Manufacturing processes includes extensive inspections to ensure that devices meet specifications; however, there is no control how the devices are handled/manipulated in the field.Therefore, the most probable conclusion code for this complaint is "cause not established" since the investigation finding do not lead to a clear conclusion about the cause of the reported event.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly, and performance specifications at the time of release for distribution.A similar complaint trend review was performed and no other complaints have been reported for the specified lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEMINI
Common Device Name
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8798966
Report Number3005099803-2019-03606
Device Sequence Number3272837
Product Code FFL
UDI-Device Identifier08714729087410
UDI-Public08714729087410
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model NumberM0063301080
Device Catalogue Number330-108
Device Lot Number0023420953
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2019
Initial Date Received by Manufacturer 06/20/2019
Supplement Date Received by Manufacturer08/08/2019
Initial Report FDA Received Date07/17/2019
Supplement Report FDA Received Date08/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-