MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-40 |
Device Problems
Pumping Stopped (1503); Battery Problem (2885)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 07/15/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare professional (hcp) via a manufacturer¿s representative (rep) regarding the patient¿s implantable drug infusion device.
The drugs being delivered were unknown.
The reason for use was spinal pain.
It was reported that a motor stall was seen at initial interrogation.
The issue occurred on (b)(6) 2019.
The patient did not recently have an mri.
Pump status and/or event log report motor stall occurred and was active.
Electromagnetic interference (emi) considerations were reviewed and it was confirmed there are no emi/magnetic sources present.
Motor stall considerations were reviewed, including to silence audible alarms, to consider pump replacement, to consider programming minimum rate, and to cover the patient with non-pump medication.
If explanted/replaced, the pump return to the manufacturer was recommended.
There were no symptoms reported.
There were no further complications reported/anticipated.
|
|