MAUDE Adverse Event Report: BOVIE MEDICAL CENTER SURGISTAT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOVIE MEDICAL CENTER SURGISTAT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number SURGII-J
Medical Device Problem Codes Inaccurate Delivery (2339); Failure to Calibrate (2440)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior the procedure, the device exceeded the output values were detected.Also, rem test failed was noted.There was no patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received as a service request and a visual inspection and functional test were performed.The returned device did not meet specification as received by the supplier.The reported condition was confirmed.According to the technician, "issue confirmed for nem out of specification.The unit meets all wi-tsu-006 rev 3 specs." the investigation found the most probable cause of the reported event to be wear and tear.The nem was recalibrated to address the issue.The unit was subjected to full factory testing / calibration with qa-327_22.It passed all tests with no problems or issues found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior the procedure, the device exceeded the output values were detected.Solid pad upper trip resistance is 10 ohm.Split pad upper trip resistance is 153 ohm.Also, return electrode monitoring(rem) test failed was noted.There was no patient involvement.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGISTAT
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOVIE MEDICAL CENTER
5115 ulmerton rd
clearwater FL 33760
MDR Report Key8800484
Report Number1717344-2019-00918
Device Sequence Number335662
Product Code GEI
UDI-Device Identifier10884521041714
UDI-Public10884521041714
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSURGII-J
Device Catalogue NumberSURGII-J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2019
Initial Date Received by Manufacturer 07/01/2019
Supplement Date Received by Manufacturer08/05/2019
10/09/2019
Initial Report FDA Received Date07/17/2019
Supplement Report FDA Received Date08/27/2019
10/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-