| Model Number |
6533 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.During a procedure, the 20mm x 2.50mm nc quantum apex balloon catheter was caught in the guidewire.The procedure was completed with a different device and no patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.During a procedure, the 20mm x 2.50mm nc quantum apex balloon catheter was caught in the guidewire.The procedure was completed with a different device and no patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an nc quantum apex balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was contrast in the inflation lumen and balloon, and blood in the guidewire lumen.The balloon was loosely folded.Microscopic inspection revealed tip damage.The device was functionally tested with a.014 guidewire.Guidewire was unable to be inserted thru the tip due to tip damage.Guidewire encountered resistance at tip when advanced thru exit notch.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|