Model Number ZCT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2019 and (b)(6) 2019.(b)(4).All pertinent information available to johnson and johnson surgical vision has been submitted.
 
Event or Problem Description
It was reported that tecnis toric intraocular lens (model zct150 +26.0 diopter) was explanted from patient¿s left eye in secondary surgical procedure because of patient experiencing ghost like image, double image and residual astigmatism.At explant there was no incision enlargement, no vitrectomy and no sutures required.A model pcb00 +25.5 diopter was used as the replacement lens for the patient.Patient has recovered.No further information provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes, returned to manufacturer on: 7/30/2019.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint cannot be confirmed.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints related to this production order was conducted in the complaint system.The search revealed that no similar complaints were received for this production order number.Product quality deficiency was not identified.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.H3 other text : placeholder.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-(b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key8800931
Report Number9614546-2019-00625
Device Sequence Number2839110
Product Code HQL
UDI-Device Identifier05050474552265
UDI-Public(01)05050474552265(17)191014
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/12/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/14/2019
Device Model NumberZCT150
Device Catalogue NumberZCT150U260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2019
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/19/2019
Supplement Date Received by Manufacturer08/02/2019
10/25/2020
Initial Report FDA Received Date07/17/2019
Supplement Report FDA Received Date08/16/2019
11/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-