• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROMUS ELEMENT PLUS; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROMUS ELEMENT PLUS; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 9394
Device Problems Fracture (1260); Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/05/2019
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Promus element plus mr ous 3.50 x 38mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent od (outer diameter) was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks along the hypotube shaft and a break situated at 22.7 cm distal end of the strain relief.The type of damages noted most likely occurred due to excessive forces that could have been applied on the delivery system.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.
 
Event Description
It was reported that crossing difficulties were encountered and fractured.The 80-90% diffuse stenosis was located in the right coronary artery.A 3.50 x 38 mm promus element plus was advanced but failed to cross the lesion and fractured.The device was completely removed from the patient.The procedure was completed with a different device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS ELEMENT PLUS
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key8802452
MDR Text Key151433306
Report Number2134265-2019-08418
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/10/2020
Device Model Number9394
Device Catalogue Number9394
Device Lot Number0022255744
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/21/2019
Initial Date FDA Received07/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
-
-