• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ZIMMON BILIARY STENT; FGE CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ZIMMON BILIARY STENT; FGE CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number G22167
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195 importer site establishment registration number: 3005580113.Device evaluation: 1 x zso-7-7 of lot number c1520348 was returned to cirl for evaluation open in its original packaging.Lab evaluation: the device involved in the complaint was evaluated in the laboratory on 12th july 2019.The stent was observed to be kinked at the 2nd, 3rd & 5th portholes on the tapered end & at the 5th porthole on the non-tapered end.The wireguide had perforated the stent at the 2nd porthole on the tapered end.A 0.035-inch wire guide would not pass the kinks.Documents review including ifu review: prior to distribution all zso-7-7 are subjected to visual and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place in cirl.A review of the manufacturing records for zso-7-7 of lot number c1520348 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1520348.The instructions for use, ifu0045-6 which accompanies this device instructs the user to ¿visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use.¿ also, the user is instructed by the instructions for use, ifu0045-6: ¿care must be exercised when straightening the pigtail curls* in order to avoid kinking or breaking the stent.¿ there is no evidence to suggest that the user did not follow the instructions for use.Root cause review: a definitive root cause cannot be determined, as the circumstances of use cannot be replicated in the laboratory and due to limited information.A possible cause of this complaint may be that the kink occurred while being straightened as it was being loaded onto the wire guide.Summary: complaint is confirmed as the failure was verified in the laboratory.According to the information reported, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
When inserting a plastic stent into the guide wire, the pigtail was kinked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMON BILIARY STENT
Common Device Name
FGE CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key8802768
Report Number3001845648-2019-00338
Device Sequence Number10848436
Product Code FGE
UDI-Device Identifier00827002221673
UDI-Public(01)00827002221673(17)210716(10)C1520348
Combination Product (Y/N)N
PMA/510(K) Number
K851962
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/16/2021
Device Model NumberG22167
Device Catalogue NumberZSO-7-7
Device Lot NumberC1520348
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/17/2019
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2019
Initial Report FDA Received Date07/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-