| Model Number |
E3773-36 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/28/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during trans abdominal pre-peritoneal, the electrode tip of the devices were broken and the part fell into the patient's cavity and was retrieved.When a new product was opened, the electrode tip was broken again.The third new product was used without problems, so the procedure was proceeded without problems.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.The investigation found the device was intact.The clear insulation near the tip was not damaged or missing.The tip of the electrode was intact and not damaged.The insulation on the shaft of the electrode was intact and not damaged.The investigation could not determine the root cause of the customer's report.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|