BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® URINE COMPLETE CUP KIT; TRANSPORT CULTURE MEDIUM Back to Search Results
Catalog Number 364956
Medical Device Problem Code Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 10/19/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.Additionally, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.Investigation conclusion: as no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Root cause description: as there was no sample or photo available for evaluation, a root cause could not be determined.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that there has been needle stick injuries while using bd vacutainer® urine complete cup kit.This was discovered after use.The following information was provided by the initial reporter: material no.364956, batch no.Unknown.It was reported that there has been 3 accidental needle stick injuries occurred with urine cup, over a course of 5 years.(2 of 3: date of event (b)(6) 2017).When i tried to run my report again the computer only came up with 3 in the last 5 years.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® URINE COMPLETE CUP KIT
Common Device Name
TRANSPORT CULTURE MEDIUM
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
MDR Report Key8804066
Report Number1917413-2019-01772
Device Sequence Number321197
Product Code JSM
UDI-Device Identifier50382903649568
UDI-Public50382903649568
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K024240
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number364956
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2019
Initial Report FDA Received Date07/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-