Model Number PED-475-18
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pipeline flex has not been returned for evaluation; product analysis cannot be performed.The device was not returned; the reported event could not be confirmed.The cause of the event could not be conclusively determined from the reported information.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report of pipeline flex delivery wire fracture.The patient was undergoing placement of the pipeline flex in the treatment of an unruptured, amorphous aneurysm in the left, paraoph thalmic internal carotid artery.Vessel tortuosity was described as severe.During the procedure, a continuous flush was maintained.The pipeline flex was reportedly placed in the intended location.At recapture of the pipeline flex delivery wire, resistance was experienced in the microcatheter.The delivery wire fractured at the point of the sodder joint just proximal to the ptfe sleeves.The delivery wire was able to be captured in the micro catheter.There were no reports of patient injury in association with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key8804166
Report Number2029214-2019-00757
Device Sequence Number13532974
Product Code OUT
UDI-Device Identifier00847536016712
UDI-Public00847536016712
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P100018.S011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/21/2022
Device Model NumberPED-475-18
Device Lot NumberA787743
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2019
Initial Report FDA Received Date07/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
-
-