| Catalog Number |
23534107 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was not returned.If additional information becomes available, it will be provided in a supplemental report.Device disposition unknown.
|
| |
|
Event or Problem Description
|
|
Field warehouse team noticed that the knob of the tibia targeting arm is cracked.The device is still functional.Not known when or under what circumstances the device cracked.
|
| |
|
Event or Problem Description
|
|
Field warehouse team noticed that the knob of the tibia targeting arm is cracked.The device is still functional.Not known when or under what circumstances the device cracked.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection was limited to the image of the device.Based on the image provided a crack could be verified at the grey tightening knob.The dhr of the reported tibia target device revealed no manufacturing discrepancies.Further investigation determined that the crack was caused by a manufacturing agent used by a contract manufacturer.This is being investigated as a nonconformity.
|
| |
|