• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH OUTER SHEATH, 26 FR., 2 STOPCOCKS, ROTATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH OUTER SHEATH, 26 FR., 2 STOPCOCKS, ROTATABLE Back to Search Results
Model Number A22026A
Health Effect - Clinical Code Tissue Damage (2104)
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer was informed by the doctor at the user facility that the connection where the hf resection inner and outer sheath meet (a22040a and a22026a) has the potential to tear the urethra of the patients.The doctor reported that the event has happened before but the doctor did not provide additional information regarding the event or patient.Additionally, the doctor reported that they believe it is one resectoscope set of sheaths that caused the reported event.The user facility utilizes 3rd party repairs for their equipment.This is 2 of 2 devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OUTER SHEATH, 26 FR., 2 STOPCOCKS, ROTATABLE
Common Device Name
OUTER SHEATH
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key8804572
Report Number2951238-2019-01017
Device Sequence Number343079
Product Code GCP
UDI-Device Identifier04042761029346
UDI-Public04042761029346
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/18/2019
Report Date (Section F) 06/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA22026A
Device Catalogue NumberA22026A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/21/2019
Event Location Hospital
Date Report to Manufacturer06/21/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INNER SHEATH:OLYMPUS A22040A LOT UNKNOWN
Outcome Attributed to Adverse Event Other;
-
-