| Catalog Number |
1550300-15 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device code 2017: excessive force the device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that a 3.0x15mm xience sierra stent delivery system (sds) was advanced over a non-abbott guide wire.The sds became stuck with the guide wire, and the sds shaft became separated during removal.A larger quantity of contrast was used than usual amount.Another non-abbott guide wire was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was returned for analysis.The reported difficult to advance and the reported difficult to remove were able to be confirmed.The reported material separation was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported difficulties appear to be related to circumstances of the procedure and subsequent deviation of the instructions for use (ifu) as it is likely that during advancement onto the guide wire interaction of devices possibly due to procedural contaminants (blood, contrast) resulted in the noted bunched inner member; thus resulting in the reported difficulty to advance and the reported difficulty to remove.It should be noted that the xience sierra everolimus eluting coronary stent system instructions for use (ifu) states: if unusual resistance is felt before the stent exits the guiding catheter, do not force passage.Resistance may indicate a problem and the use of excessive force may result in stent damage or dislodgement.Manipulation of the device using force resulted in the noted device damages (stretched inner/outer member, bunched balloon) and ultimately resulted in the reported material separation/noted inner and outer member separations.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|