Extremely premature infant delivered at (b)(6), intubated in the delivery room and admitted to nicu.A 3.5fr double-lumen umbilical venous catheter (uvc) was placed.Following an x-ray to determine appropriate placement, the neonatal nurse practitioner (nnp) gave an order to pull the uvc back 0.5cm.When the rn gently pulled back on the uvc, it reportedly snapped in half.The portion remaining in the infant was held until the nnp could successfully remove it.Staff reported the catheter broke below the suture line.Insertion distance - 6.Line was secured by suture.Nicu manager reported the break was below the suture line and didn't appear to be caused by the suture.
|