Model Number 8888541023P
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during use, the nurse saw a leak at the venous extension tube which is located at the junction of extension tube and bifurcate.It was stated that the catheter was repaired by a new material from a kit with unknown lot number to resolve the issue.The treatment was proceeded after the replacement of catheter and the treatment was completed.There was blood loss of less that 1(one) ml(milliliter).Tego was not utilized and there was no luer adapter issue.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led photographic evaluation of one device.The photo depicts the blue and red luer adapters, the clamps and the hub in a small bag.The cannula is not attached.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the disengaged cannula may occur when mishandled during clinical application.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME
Common Device Name
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key8807760
Report Number3009211636-2019-00172
Device Sequence Number326965
Product Code MSD
UDI-Device Identifier10884521158504
UDI-Public10884521158504
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K123196
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model Number8888541023P
Device Catalogue Number8888541023P
Device Lot Number1807300167
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/02/2019
Supplement Date Received by Manufacturer08/21/2019
Initial Report FDA Received Date07/19/2019
Supplement Report FDA Received Date09/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age14 YR
-
-