Catalog Number 4401206S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, a powerflex pro pta balloon catheter (12mm6cm 80) ruptured at 12-atmospheres pressure (atm).The balloon popped and came out fully intact.There was no reported patient injury.Initially, this was a fistulagram case.The device will not be returned for analysis as the device was discarded.
 
Additional Manufacturer Narrative
A powerflex pro percutaneous transluminal angioplasty (pta) balloon catheter (12mm x 6cm x 80cm) ruptured at 12-atmospheres (atm).The balloon popped and was removed intact from the patient.There was no reported patient injury.Initially, this was a fistulagram case.The product was not returned for analysis.A product history record (phr) review of lot 17690869 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst-at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.With the paucity of information available and without return of the product for analysis it is not possible to draw a clinical conclusion between the device and the reported event.However, vessel characteristics and procedural factors may have contributed to the reported event.According to the safety information in the instructions for use ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 12MM6CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014
MDR Report Key8807956
Report Number9616099-2019-03089
Device Sequence Number326815
Product Code LIT
UDI-Device Identifier20705032060193
UDI-Public20705032060193
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2020
Device Catalogue Number4401206S
Device Lot Number17690869
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2019
Supplement Date Received by Manufacturer07/21/2019
Initial Report FDA Received Date07/19/2019
Supplement Report FDA Received Date08/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-