The clamp device is essential to occlude the aorta and provide the necessary cardiac isolation required to perform minimally invasive cardiac surgery procedures.If the balloon bursts during a procedure, the heart would fill and warm, the operative site may be obscured and the procedure may need to convert to an open procedure.In this case, all fragments were able to be retrieved the patient was reported to be doing fine.Follow up for the device is still in progress and the investigation is pending.A supplemental report will be submitted upon receipt of additional information.
|
As received, the device was returned with visible traces of blood and was examined in the biohazard area of the lab.Balloon was observed to be ruptured (see figure 2).Edges of rupture site appeared to match up.All through lumens were found to be patent without any leakage or occlusion.No other visual damage, contamination, or other abnormalities were found.Per the product evaluation, the customer complaint of balloon rupture was confirmed.Per the engineering evaluation, a supplier manufacturing defect was confirmed.A design, ifu, and labeling defect were not confirmed.The trend is in control and the fmea line item is appropriate.A pra was initiated and the root cause will be investigated in the capa.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|