• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP; STEM, HUMERAL SIZE6 RSP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP; STEM, HUMERAL SIZE6 RSP Back to Search Results
Model Number 506-00-006
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/20/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was due to peri-prosthetic fracture.The previous surgery and the revision detailed in this investigation occurred over 7 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.Initial or prolonged hospitalization was required.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports associated with the product that may have contributed to the event.The device and its applicable concomitant devices were within their respective expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to peri-prosthetic fracture.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's event.There are multiple factors that may contribute to the event that are outside the control of djo surgical are excessive loading, poor bone density, lack of care, patient activities or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient had a peri-prosthetic fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Common Device Name
STEM, HUMERAL SIZE6 RSP
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8809404
Report Number1644408-2019-00673
Device Sequence Number2754949
Product Code PHX
UDI-Device Identifier00888912024372
UDI-Public(01)00888912024372
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K041066
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2017
Device Model Number506-00-006
Device Catalogue Number506-00-006
Device Lot Number810C1045
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2019
Initial Report FDA Received Date07/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
-
-