Model Number 20-006-021
Health Effect - Clinical Codes Cardiac Arrest (1762); Death (1802); Unspecified Infection (1930)
Date of Event 06/10/2019
Type of Reportable Event Death
Additional Manufacturer Narrative
There is no indication or allegation from the field/clinic that the device contributed to the incident.Luminex's investigation has determined that this is not a device malfunction.The product is adequately labeled and the product performed according to product claims.Luminex has made multiple requests for additional information from the medical director with regards to this case, but has not received any more information.Luminex's investigation was not able to determine root cause as the patient sample was discarded and not available for internal testing.Instrument and consumable data was reviewed and documentation showed no indication of device or consumable malfunction.The package insert labeling clearly states in rare instances for specimens with organisms carrying a resistance marker, bc-gn may not yield a positive result for the resistance marker when the organism(s) is detected.The package insert labeling states: bc-gn is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, is not used to monitor these infections.The package insert labeling states: a not detected result for any of the antimicrobial resistance markers does not indicate susceptibility, as resistance may occur by other mechanisms.Sub-culturing and ast testing is required in order to assign a resistant and/or susceptible phenotype to each isolate recovered from the blood culture sample.
 
Event or Problem Description
On (b)(6) 2019 sample (b)(6) was processed using the cartridge id (b)(6) verigene gram-negative blood culture test (bc-gn) on the verigene system.Verigene result was pseudomonas aeruginosa detected and reported out to a physician on (b)(6) 2019.Patient sample was run on vitek, and the vitek result was reported on (b)(6) 2019 as p.Aeruginosa identified, cefazolin and meropenem resistant.The hospital facility notified luminex (nanosphere) of the discrepancy between the bc-gn and vitek results on (b)(6) 2019 and provided details of the treatment.The patient was treated with meropenem and tobramycin for 48 hours following the verigene result.Following the vitek result, treatment was altered and meropenem was discontinued on (b)(6) 2019.Ciprofloxacin and avycaz were added to treatment on (b)(6 2019.On (b)(6) 2019, the patient expired from cardiac arrest.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERIGENE GRAM-NEGATIVE BLOOD CULTURE TEST (BC-GN)
Common Device Name
GRAM-NEGATIVE BLOOD CULTURE TEST
Manufacturer (Section D)
NANOSPHERE, INC.
4088 commercial avenue
northbrook IL 60062
MDR Report Key8809480
Report Number3006028115-2019-00001
Device Sequence Number19619795
Product Code PEN
UDI-Device Identifier00857573006140
UDI-Public00857573006140
Combination Product (Y/N)N
PMA/510(K) Number
K132843
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/13/2019
Device Model Number20-006-021
Device Catalogue Number20-005-021
Device Lot Number031519021B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2019
Initial Report FDA Received Date07/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age62 YR
-
-