PENUMBRA, INC. PENUMBRA SYSTEM 3D REVASCULARIZATION DEVICE; NRY Back to Search Results
Catalog Number PSR3D
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Rupture (2208)
Date of Event 06/20/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for evaluation.From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported event.Potential adverse events in the labeling with the penumbra system include, but are not limited to, hematoma or hemorrhage at the site, inability to completely remove thrombus, intracranial hemorrhage, ischemia, including death.Therefore, it was determined that the reported adverse events were anticipated complications.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2019-01412.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the left m2 segment of the middle cerebral artery (mca) using a penumbra system jet7 kit (kit) and a penumbra system 3d revascularization device (psr3d).During the procedure, the physician made two successful passes using a penumbra system jet 7 reperfusion catheter (jet7) with psr3d, a velocity delivery microcatheter (velocity) and a non-penumbra guidewire.After making a third and final pass with the jet7 and psr3d; the physician noticed a rupture upon removing the jet7 and psr3d.Therefore, the physician advanced a non-penumbra balloon catheter and a guidewire; however, it was not inflated.It was discovered under pre-inflation run that the rupture had sealed itself off and closed.The procedure had ended at this point.The physician noted that the rupture may be due to patient¿s age and anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM 3D REVASCULARIZATION DEVICE
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8809608
Report Number3005168196-2019-01413
Device Sequence Number18912252
Product Code NRY
UDI-Device Identifier00814548017839
UDI-Public00814548017839
Combination Product (Y/N)Y
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K162901
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/20/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Catalogue NumberPSR3D
Device Lot NumberF88327
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2019
Initial Report FDA Received Date07/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Date Report Sent to FDA01/01/2005
-
-