| Model Number |
7060M-100 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
07/05/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmea, the most probable root cause is user error; improper implant size selection, using too long of an implant or installing the implant too deep.
|
| |
|
Event or Problem Description
|
|
The patient had right side si joint arthrodesis in (b)(6) 2019 where three implants were installed.The patient complained of radicular pain following the initial procedure.The surgeon determined that the middle-positioned implant had breached the neuroforamen and was impinging on the nerve root.The day after the initial procedure, the surgeon performed a revision procedure where he added an additional implant of the same type in a more dorsal position and then removed the middle-positioned implant.No other preexisting implants were adjusted or removed.The patient's radicular pain symptoms resolved following the revision procedure.
|
| |
|