• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS SYMBOTEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS SYMBOTEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number SYM2015OS
Medical Device Problem Codes Mechanics Altered (2984); Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Abscess (1690); Adhesion(s) (1695); Emotional Changes (1831); Fistula (1862); Unspecified Infection (1930); Pain (1994); Seroma (2069); Injury (2348); Impaired Healing (2378); Ascites (2596); Bowel Perforation (2668); No Code Available (3191); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Death
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a hernia.It was reported that after implant, the patient experienced non-healing wound, abscess, infection, bowel perforation, fistula, leukocytosis and seroma.Post-operative patient treatment included bowel resection and revision surgery.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a ventral hernia.It was reported that after implant, the patient experienced non-healing wound, abscess, infection, bowel perforation, fistula, leukocytosis, adhesions, large amount of bilious fluid material and succus in abdomen, and seroma.Post-operative patient treatment included bowel resection/transected, transferred to icu, mesh removal, and revision surgery.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a ventral hernia.It was reported that after implant, the patient experienced non-healing wound, abscess, infection, bowel perforation, fistula, leukocytosis, adhesions, large amount of bilious fluid material and succus in abdomen, seroma, defective device, mental pain, disability, impairment, loss of enjoyment of life, and pain.Post-operative patient treatment included bowel resection/transected, transferred to icu, mesh removal, and revision surgery.
 
Additional Manufacturer Narrative
Additional information: b2 (disability), b5, d8, e1 (facility name, street 1, city, region, postal code), g1 (mfr contact first name, last name, street 1, mfr city, region, postal code, email, phone number), g3, g4(510k), h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a ventral hernia.It was reported that after implant, the patient experienced non-healing wound, abscess, infection, bowel perforation, fistula, leukocytosis, adhesions, large amount of bilious fluid material and succus in abdomen, seroma, defective device, mental pain, disability, impairment, loss of enjoyment of life, small bowel enterotomies, pain, and death.Post-operative patient treatment included bowel resection/transected, transferred to icu, mesh removal, drained fluid from abdomen, ileostomy closure, physical therapy, lysis of adhesions, transected fistula, hospitalization, and revision surgery.Information received indicates the patient is deceased.No information was provided regarding the circumstances of expiration.
 
Additional Manufacturer Narrative
Additional information: b2, b5, h1, h6 (added imf codes) complaint reportability has been reassessed and is now reportable as a death.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMBOTEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
5920 longbow drive
8200 coral sea st ne
mounds view,mn, CT 55112
7635265677
MDR Report Key8810106
Report Number9615742-2019-02610
Device Sequence Number2838528
Product Code FTL
UDI-Device Identifier10884521190467
UDI-Public10884521190467
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K142908
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model NumberSYM2015OS
Device Catalogue NumberSYM2015OS
Device Lot NumberPOK0386X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/03/2019
Supplement Date Received by Manufacturer03/04/2020
07/13/2022
01/18/2023
Initial Report FDA Received Date07/21/2019
Supplement Report FDA Received Date03/26/2020
07/28/2022
01/27/2023
Date Device Manufactured11/26/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN PROTACK (LOT# UNKNOWN).
Outcome Attributed to Adverse Event Death; Hospitalization; Required Intervention; Disability;
Patient SexMale
-
-