COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIG30AV
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a lobectomy procedure, the surgeon tried to fire, but the handle could not be squeezed and the device could not be fired.The procedure was completed with another device.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the reload was pre-fired and engaged in interlock.There were staples protruding from proximal staple cartridge channel.Functionally the reload was loaded into a post market vigilance instrument, the interlock was overridden, and the reload was applied to test media with proper staple placement and media transection.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the pre-fire condition with interlock engagement may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.In this situation, the safety interlock feature will engage and prevent the reload from firing a second time by ceasing the placement of staples and tissue transection and prevent patient harm.The root cause of the observed damage was due to the product not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8810125
Report Number1219930-2019-04264
Device Sequence Number3026749
Product Code GDW
UDI-Device Identifier20884521543673
UDI-Public20884521543673
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberSIG30AV
Device Catalogue NumberSIG30AV
Device Lot NumberN8L0785Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2019
Initial Date Received by Manufacturer 07/03/2019
Supplement Date Received by Manufacturer08/19/2019
Initial Report FDA Received Date07/21/2019
Supplement Report FDA Received Date09/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-