DAVIS & GECK CARIBE LTD VERSAPORT PLUS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 179095P
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, preoperatively, the obturator/trocar broke.Another device was used to complete the case.There was no patient involvement.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted: the product received was un-open in its appropriate packaging (blister).In addition, it was observed that besides the loose spring in package, the obturator cap was open, moreover, the cover of the obturator had a missing tab.The broken tab of the cover was not received in the sealed package.Functional testing of the trocar was precluded due to the received condition.A review of the device history record for the trocar indicates the product was released meeting all medtronic quality release specifications at the time of manufacture.The root cause of the damaged cap was determined to be a result of a component obtained from a supplier.A process improvement has been initiated to prevent this condition from recurring.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPORT PLUS
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 18750
DO  18750
MDR Report Key8810355
Report Number9612501-2019-01422
Device Sequence Number2824053
Product Code GCJ
UDI-Device Identifier10884521081284
UDI-Public10884521081284
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K952977
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number179095P
Device Catalogue Number179095P
Device Lot NumberJ8B1412X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2019
Supplement Date Received by Manufacturer09/16/2019
11/15/2019
Initial Report FDA Received Date07/22/2019
Supplement Report FDA Received Date10/08/2019
12/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-