| Catalog Number |
304 |
| Medical Device Problem Code |
Labelling, Instructions for Use or Training Problem (1318)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the system determined that the system was operating as intended at the time of the incident.Per discussion with the personnel involved, the bedside user lifted the guidewire above the axis of rotation during a guidewire rotation move.As such, this caused the guidewire to dislodge from the drive track.This error has the possibility of causing damage to the guidewire.The users have been retrained by corindus clinical personnel as to the proper handling techniques for guidewire management.
|
| |
|
Event or Problem Description
|
|
Per the facility, during robotic delivery of the guidewire, the wire wrapped around the rear of the rotary drive of the cassette.
|
| |
|