• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number RLT351416
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/21/2019
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2019, this patient underwent endovascular repair to treat an abdominal aortic dissection with an internal iliac artery aneurysm.The patient was reported to have poor anatomy and poor aortic tissue prior to the procedure.The internal iliac artery aneurysm was coil embolized.Then a gore® excluder® aaa endoprosthesis featuring c3 deployment system was advanced into the abdominal aorta.When the device was deployed, the contralateral leg gate opened into a dissected plane.Therefore, the physician was unable to cannulate the gate.The decision was made to convert to an open repair.The gore® excluder® device was explanted and discarded at the facility.There were no allegations against the device.The physician stated that the patient's anatomy was poor, and this was the cause of the conversion to open repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key8811736
Report Number3007284313-2019-00209
Device Sequence Number2587351
Product Code MIH
UDI-Device Identifier00733132618484
UDI-Public00733132618484
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/24/2022
Device Catalogue NumberRLT351416
Device Lot Number20390956
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
-
-